Clean, package, load and run a tabletop steam sterilizer, read its indicators and spore tests, and know what to do when a check fails.

Table of contents

Suture scissors, forceps and needle holders go into a patient's skin or a wound. If one is not sterile, it can carry germs from the last patient into the next. In many clinics, the medical assistant runs the steam sterilizer, often called an autoclave, and has to do it the same careful way every time.

Three sources set the rules. The CDC's Guideline for Disinfection and Sterilization in Healthcare Facilities gives the core practices. AAMI ST79 is the national standard for steam sterilization in health care facilities, and your employer's procedure may be based on it. The manufacturer's instructions for use (IFU) for your sterilizer and for each instrument give the cycles, loading and care. When the sterilizer's and an instrument's instructions disagree, CDC says to resolve it with both manufacturers, not to guess.

You need hands-on training before you run a load on your own. CDC's outpatient guidance says staff who reprocess instruments should show competency, with a trainer watching, when first assigned, every year, and whenever new devices or procedures arrive. Reading prepares you for that training. It does not replace it.

What You Will Be Able to Do

  • Explain how steam sterilizes, and why soil and trapped air cause failures.
  • Clean, inspect and package instruments for a steam cycle.
  • Load the sterilizer, pick the cycle, and unload dry packs safely.
  • Read mechanical, chemical and biological indicators and keep the log.
  • Respond correctly to a failed indicator or a positive spore test.

Which Items Get Sterilized

Anything that enters sterile tissue or the bloodstream must be sterile. In a medical office, that means instruments for suturing, suture removal, biopsies, incision and drainage, and minor surgery. Clean every item first, then match the processing to how it touches the body.

Two kinds of items never go in the autoclave. One is anything labeled single use, marked with a circled 2 crossed by a line. Throw it away. The other is anything whose instructions do not list steam as an approved method.

How Steam Sterilizes

CDC names four parameters of steam sterilization: steam, pressure, temperature and time. Steam under pressure gets hotter than boiling water and kills germs, including bacterial spores, but only where it touches. Three things block that contact:

  • Soil. Dried blood or tissue shields the germs underneath it.
  • Trapped air. A cool air pocket in a crowded chamber or a closed instrument can stay below sterilizing temperature.
  • Wrong cycle. A wrapped load needs a longer exposure than unwrapped instruments.

CDC gives two common minimum exposures: 30 minutes at 121°C (250°F) in a gravity displacement sterilizer, and 4 minutes at 132°C (270°F) in a prevacuum sterilizer. These are times at temperature, not the whole cycle, since heat-up and drying add more. The exact cycle for each kind of load comes from your sterilizer's instructions.

Type How it removes air Extra test
Gravity displacement Steam pushes air out through a vent None beyond the standard monitors
Prevacuum A pump pulls air out before steam enters Air removal test each day it is used

Many tabletop units are gravity displacement. A prevacuum unit needs an air removal test, often called a Bowie-Dick test, run in an empty chamber each day the sterilizer is used. If it fails, CDC says not to use the sterilizer until it is inspected and passes.

Step 1: Contain and Transport

Work moves one way, from the dirty sink to packaging, the sterilizer and storage, and dirty and clean areas stay apart. Treat every instrument as contaminated until it has been through a full cycle.

  1. Remove disposable sharps at the point of use. Blades and needles go into the sharps container in the exam room.
  2. Contain reusable instruments right away. OSHA's bloodborne pathogens standard requires contaminated reusable sharps to go in a container that is puncture resistant, leakproof on the sides and bottom, and labeled or color-coded.
  3. Keep soil from drying. CDC says to clean instruments as soon as practical after use. If you cannot clean them soon, soak them in your clinic's detergent or enzymatic cleaner. A high-level disinfectant such as glutaraldehyde is not a holding solution.

Warning: Never reach into a basin of soapy water to feel for instruments. CDC's dental guidance says to hold them in a strainer-type basket and lift them out with forceps.

Step 2: Clean

Wear puncture-resistant, heavy-duty utility gloves, which CDC recommends for cleaning contaminated instruments, plus a mask, eye protection or face shield, and a gown, because splashing is likely. OSHA lets you reuse utility gloves after decontaminating them, but throw them away once they are cracked, peeling, torn or punctured.

  1. Mix the cleaner the instrument instructions call for, at the strength on its label. CDC's outpatient checklist says to discard it as its maker directs, typically after each use.
  2. Open every hinge. Box locks and serrated jaws hold debris.
  3. Clean manually or mechanically. By hand, scrub below the waterline to limit splashing, with a long-handled brush that keeps your hand away from points and edges. With an ultrasonic cleaner, follow its manual for solution, fill level and run time, and keep the lid closed while it runs. Machine cleaning means less handling of sharp, dirty instruments.
  4. Rinse off all detergent, which can leave residue and stains.
  5. Dry the instruments before you package them.

Step 3: Inspect and Package

Under good light, look for soil in every box lock, serration and hinge, and clean again if you find any. Check that scissors cut, jaws line up and ratchets hold, with no cracks, rust or pitting. Take a damaged instrument out of service and tell your supervisor.

  1. Use packaging made for steam. CDC says it must be compatible with the process and cleared by the FDA. Peel pouches and sterilization wrap are common.
  2. Keep hinged instruments open and unlocked, so steam reaches where the jaws meet.
  3. Take instruments apart if their instructions say to.
  4. Put a chemical indicator inside every package. It changes color when steam reaches the inside. If you cannot see it from outside, also use an external indicator, such as indicator tape.
  5. Do not overfill. Contents should lie flat with room to spare, and the seal must be closed along its whole length.
  6. Label each package with the sterilizer, the load or cycle number, the date, and an expiration date if your clinic uses one. Write where the pouch maker directs, usually the label area or clear side, not the paper.

Step 4: Load the Sterilizer

  1. Check the water reservoir. Fill it only with the type of water the manual names, such as distilled water.
  2. Follow the loading diagram in the manual. CDC says to place items correctly and loosely, so steam can circulate. Overloading is a known cause of failed loads.
  3. Stand peel pouches on edge in a rack, as CDC recommends, with space between them. Do not stack them flat.
  4. Pick the cycle that matches the load, such as unwrapped, pouches or packs. A cycle for unwrapped instruments is not made for wrapped ones.

An unwrapped load is for an item needed right away, sometimes called immediate-use steam sterilization. CDC says it is not for convenience, and the items are used at once, not stored. Your clinic's policy says when, if ever, you may do this.

Step 5: Run, Dry and Unload

  1. Read the mechanical record when the cycle ends. Check that the time, temperature and pressure met the cycle's settings on the display or printout. CDC says not to use the items if they did not.
  2. Let the packs dry inside the chamber. CDC says not to touch packs until they are cool and dry, because hot packs act as wicks and pull in moisture, and bacteria with it, from your hands.
  3. Check every indicator as you unload. If an internal indicator did not change, do not use that package.
  4. Look for wet packs. CDC says sterile items that become wet are considered contaminated. Reprocess them from the start.

Warning: Steam and hot metal cause serious burns. Open the door only when the cycle has ended and the manual says it is safe. Stand to the side as you open it, and use heat-resistant gloves for hot trays.

Step 6: Store

Store sterile packages in closed or covered cabinets, in a clean, dry area away from sinks. Under event-related shelf life, which CDC describes, a package stays sterile until something compromises it. Check every package before you open it. If it is torn, wet, punctured or the seal is open, reprocess it. Use older packages first.

The Three Kinds of Monitoring

No single check proves a load is sterile, so CDC calls for all three.

Monitor What it tells you How often
Mechanical Time, temperature and pressure reached Every load
Chemical Steam reached the package, inside and out Every package
Biological (spore test) The cycle killed highly resistant spores At least weekly, and every load with an implantable item

A biological indicator (BI) holds bacterial spores that are hard to kill. For steam, CDC names spores of Geobacillus stearothermophilus. If the cycle worked, the spores die and nothing grows when the BI is incubated. Chemical indicators do not prove sterility, and CDC says they should not replace spore tests.

Running a Spore Test

  1. Use a BI made for your type of sterilizer and cycle.
  2. Place it where the BI and sterilizer instructions say, inside a normal load, and run the cycle as usual.
  3. Keep a control from the same lot number that does not go through the cycle, as CDC's dental guidance directs. The control should grow, which proves the spores were alive. The test should not.
  4. Read the result at the time the BI's instructions give, in an office incubator or through a mail-in testing service.
  5. Record the result in the log.

A load with an implantable item gets its own BI, and CDC says to hold those items, whenever possible, until the result is negative.

When a Check Fails

Problem What it means What you do
Mechanical record off The cycle did not reach its settings Do not use the load, tell your supervisor, reprocess
Inside indicator unchanged Steam did not reach that package Do not use it, check the rest of the load, repackage and rerun
Wet pack Considered contaminated Reprocess, and report it if it keeps happening
Air removal test failed Air is trapped in a prevacuum chamber Do not use the sterilizer until it is inspected and passes

For a positive spore test, CDC's dental guidance gives these steps:

  1. Take the sterilizer out of service and review how loads were packaged, loaded and run.
  2. Repeat the spore test right away, using the same cycle that failed.
  3. Recall and quarantine any implantable items from that sterilizer until they are shown to be sterile.
  4. If the repeat test is negative and the procedures were correct, the sterilizer can go back into service.
  5. If the repeat test is positive, keep it out of service until it has been checked and has passed spore tests in three fully loaded cycles in a row. Recall and reprocess items from the suspect loads where possible.

Your clinic's procedure says who decides each step. Your job is to stop and report, never to quietly rerun a failed load. CDC says qualified personnel do repairs and preventive maintenance, following the manufacturer's instructions.

Keep the Log

For each cycle, CDC says to record the sterilizer and cycle used, the load number and contents, and the mechanical, chemical and biological results. Add your initials and the date, and keep records as long as state and local rules require. The log ties every package to a load, so a failed spore test tells you what to recall.

Common Mistakes

Mistake What it costs
Sterilizing an instrument with dried soil on it Germs survive under the soil
Closing hinged instruments Steam misses the surfaces inside the jaws
Overloading or stacking pouches flat Air pockets and failed loads
Unloading hot, damp packs Packs wick in germs and must be redone
Skipping the weekly spore test A failing sterilizer goes unnoticed

Check Your Understanding

1. A provider hands you a needle holder after a laceration repair, and you will not get to the dirty sink for an hour. What do you do?

Put it in a puncture-resistant, leakproof, labeled container and soak it in your clinic's detergent or enzymatic cleaner so the blood does not dry. Do not use a high-level disinfectant as a holding solution.

2. Two pouches feel damp when you unload, though the mechanical record looks fine. Can you put them away?

No. Wet sterile items are considered contaminated. Reprocess them, and report it if wet packs keep happening.

3. The mail-in service calls to say this week's spore test grew. What are your first two steps?

Take the sterilizer out of service and tell your supervisor, then repeat the spore test right away using the same cycle. Implantable items from that sterilizer are recalled and quarantined until shown to be sterile.

Key Takeaways

  • Steam sterilizes only where it touches, so clean first, open hinges and load loosely.
  • The sterilizer's and instruments' instructions set the cycle, loading and care, with CDC guidance and AAMI ST79 behind your employer's procedure.
  • Check the mechanical record every load and the chemical indicator in every package.
  • Run a spore test at least weekly, and with every load that has an implantable item.
  • Wet packs, unchanged indicators and failed tests mean stop, report and reprocess.