Clinical Medical Assisting

Recording a 12-Lead EKG

20 min read Training Guide

Prepare the patient and skin, place all 10 electrodes on their landmarks, record a clean tracing, fix artifact and swapped wires, and get the tracing to the provider.

Table of contents

An electrocardiogram (EKG or ECG) records the electrical activity of the heart. A 12-lead EKG uses 10 electrodes to show the heart from 12 angles, called leads. Your job is a clean tracing from electrodes in the right places. The provider reads it. One electrode a single rib space too high can make a healthy heart look like it has had a heart attack, so placement is the skill that matters.

Recording EKGs is a common medical assistant (MA) task, but your state's law and your employer's policy decide whether you may do it. Before you record one on a patient, you need hands-on practice on your clinic's machine, supervised tracings and a documented competency check. Nothing here certifies you. You never interpret an EKG or tell a patient what it shows.

What You Will Be Able to Do

  • Set up the machine and prepare the patient and the skin.
  • Find the landmarks and place all 10 electrodes.
  • Record a standard tracing and check it before the patient leaves.
  • Fix artifact and swapped lead wires.
  • Know when a tracing goes to the provider right now.

How 10 Electrodes Make 12 Leads

Four electrodes go on the limbs: right arm (RA), left arm (LA), right leg (RL) and left leg (LL). The right leg electrode is a reference that helps cut electrical noise. The machine combines the limb electrodes into six limb leads, named I, II, III, aVR, aVL and aVF.

Six electrodes go on the chest, V1 through V6, and each makes one chest lead. Because the machine combines electrodes, one misplaced or swapped electrode changes several leads at once.

Before the Patient Comes In

  1. Check the machine. Paper is loaded, the battery is charged or the machine is plugged in, and the cables have no cracks or bare wires.
  2. Check the electrodes. Use an in-date, sealed package. Dried-out gel causes a noisy tracing.
  3. Check the settings. The standard is a paper speed of 25 mm per second and a gain of 10 mm per millivolt (mV). Leave the filters at your clinic's default. The last user may have changed something, so look every time.
  4. Enter the patient's details, including name, date of birth and sex. The machine's computer statement uses age and sex, and a tracing under the wrong name is worse than none.

Preparing the Patient

  1. Identify the patient with two identifiers, such as full name and date of birth, and confirm the order.
  2. Explain the test. It is painless, no electricity goes into the body, and it takes a few minutes of lying still.
  3. Protect privacy. Ask the patient to undress from the waist up and put on a gown that opens in the front. Bare skin is needed at the wrists or arms and at the ankles or lower legs, so stockings come off. Keep the patient covered except where you are working.
  4. Position the patient lying flat on the back with a pillow under the head, arms resting at the sides and legs uncrossed. Keep hands and feet off metal rails.
  5. Keep them warm. Shivering shows up as noise on the tracing.

If the patient cannot lie flat, raise the head only as much as needed and write the position on the tracing. The American Heart Association (AHA) says a tracing taken sitting up should not be treated as equal to one taken lying flat.

Preparing the Skin

The AHA notes that cleaning and gentle abrasion of the skin reduce noise. Follow your clinic's procedure:

  • Remove lotion, oil and sweat where each electrode will go, and let the skin dry completely.
  • Clip hair that keeps an electrode from lying flat.
  • Rub the spot lightly with the prep pad or dry gauze if your procedure calls for it.
  • Press each electrode around its whole edge so it seals to the skin.

Placing the Limb Electrodes

For a standard resting EKG, the AHA places the four limb electrodes on the arms and legs, below the shoulders and hips. Common spots are the inner wrists or upper arms and the inner ankles or lower legs. Choose a fleshy spot rather than bone, and put each pair at the same level on both sides.

Do not put limb electrodes on the chest or belly for a resting EKG. That torso placement is used for exercise tests and monitoring, and it changes the tracing. If the provider asks for it, or a patient's limb cannot be used, label the tracing with what you did, as the AHA requires for torso placement.

Electrode Where US (AHA) color
RA Right arm White
LA Left arm Black
RL Right leg Green
LL Left leg Red

Many MAs remember the US colors as "white on the right, snow over grass" for the right side and "smoke over fire" for black above red on the left. Machines built to the international (IEC) standard use different colors, so trust the letters printed on each wire over the colors.

Placing the Chest Electrodes

Finding the Fourth Intercostal Space

An intercostal space is the soft gap between two ribs. Find it by touch, not by eye:

  1. Find the notch at the top of the breastbone (sternum). Slide your finger down until you feel a ridge across the bone. This is the sternal angle.
  2. Slide your finger to the patient's right of the ridge. The rib there is the second rib, and the space just below it is the second intercostal space.
  3. Walk your finger down two more spaces to the fourth intercostal space.

The Six Positions

These are the AHA positions:

Electrode Position AHA color
V1 Fourth intercostal space, right edge of the sternum Red
V2 Fourth intercostal space, left edge of the sternum Yellow
V3 Midway between V2 and V4 Green
V4 Fifth intercostal space, in line with the middle of the left collarbone Blue
V5 Level with V4, at the front fold of the armpit, or midway between V4 and V6 Orange
V6 Level with V4, in line with the middle of the armpit Purple

Place V1 and V2 first, then V4, then V3 between them, then V6, then V5. V5 and V6 go straight across at the height of V4. Do not follow the curve of the rib space up toward the armpit. The AHA calls the horizontal line through V4 the preferred placement.

Warning: A common error is placing V1 and V2 too high, in the second or third intercostal space. The AHA notes this can make the tracing look like a heart attack in the front of the heart. Count the rib spaces every time.

Placing V5 and V6 too low, in the sixth intercostal space or below, changes the wave sizes the provider uses to judge heart enlargement.

Breast Tissue

For women, the AHA recommends placing the electrodes on the chest wall under the breast. Explain what you are doing before you do it. Ask the patient to lift the breast, or lift it with the back of your hand, and place the electrode on the chest wall at the correct landmark.

Recording the Tracing

  1. Connect each wire to its electrode, matching the label on the wire. Let the wires lie without pulling on the electrodes.
  2. Ask the patient to lie still, breathe normally and not talk.
  3. Watch the screen for a few seconds until all 12 leads show a steady line.
  4. Record. A standard tracing covers 10 seconds, often printed as four columns of 2.5 seconds each, with a rhythm strip along the bottom.
  5. Check the tracing before you remove the electrodes, using the checklist below.
  6. Re-record if anything is wrong. Keep the electrodes on until the provider has the tracing, in case a repeat is needed.
  7. Remove the electrodes gently, wipe off the gel and help the patient dress.
  8. Clean the lead wires and cables as the manufacturer directs before the next patient.

Checking the Tracing

  • The name and date of birth on the tracing match the patient.
  • All 12 leads show waves, and none is a flat line.
  • The speed and gain printed on the tracing read 25 mm per second and 10 mm per mV.
  • The baseline between beats is steady and smooth.
  • In lead I, the waves normally point mostly upward. If lead I is upside down and aVR points up, check the arm wires. If they are right, tell the provider.

Reading the Grid

EKG paper has small boxes 1 mm wide and large boxes made of five small ones. At 25 mm per second:

Distance Time
1 small box 0.04 seconds
1 large box 0.2 seconds
5 large boxes 1 second

At the standard gain, 10 mm of height is 1 mV, so the calibration mark at the start of each line is two large boxes tall.

Worked example. The machine prints a heart rate, and you can cross-check it. A 10-second rhythm strip shows 12 heartbeats (QRS complexes, the tall spikes). 12 x 6 = 72 beats a minute. If the patient's pulse was 70 when you took vital signs, the numbers agree. If they disagree badly, tell the provider. Do not decide what it means.

Fixing Artifact

Artifact is anything on the tracing that does not come from the heart.

What you see Likely cause Fix
Jagged, uneven fuzz Muscle tension, cold, talking, tremor Warm blanket, support the arms, ask the patient to relax
Thick, even fuzz Electrical interference Move cords and devices away, check cables, keep skin off metal
Baseline wanders up and down Breathing, loose electrodes, lotion Re-prep the skin, press the electrodes, ask for normal breathing
Flat line in one or more leads Loose electrode or wire Find the electrode feeding that lead and reattach it
Lead I upside down Arm wires swapped Swap RA and LA back
Lead II nearly flat RA and RL wires swapped Swap RA and RL back

The AHA notes that swapping the left arm and left leg wires is harder to spot. Checking every wire label before you record is the only reliable fix.

Getting the Tracing to the Provider

Every tracing goes to the provider for review. Do not file it in the chart unreviewed.

Hand it to the provider right now, before you remove the electrodes and before the patient leaves, if:

  • The patient has chest pain or pressure, shortness of breath, dizziness or fainting, or a racing or pounding heartbeat, before or during the test.
  • The machine prints a statement about a heart attack, ischemia or a dangerous rhythm.
  • The patient looks unwell to you.

Write any symptom on the tracing with the time, such as "chest pressure during recording." If the patient becomes unresponsive, call for help and start your clinic's emergency protocol.

Warning: Never read the machine's computer statement to the patient, and never say the tracing "looks fine." A computer statement is not a diagnosis, and only the provider interprets the EKG.

Special Situations

Situation What to do
Amputation, cast or wound on a limb Place that electrode where your procedure says, and note it on the tracing
Tremor Support the limbs, and move limb electrodes up to the upper arms and thighs
Pacemaker Note it on the order and tracing, and use the pacemaker setting if the machine has one
Right-sided or back leads Record only when the provider orders them, and label the tracing

Documenting the EKG

Record the date and time, the patient's position if they were not flat, any changed electrode placement, any symptoms during the recording, and when and to whom you gave the tracing.

Check Your Understanding

1. A tracing shows lead I upside down, and aVR points upward. What do you check first?

The arm wires. Swapped RA and LA wires flip lead I. If the wires are correct, tell the provider and let them decide.

2. Partway through the recording, the patient says, "My chest feels tight." What do you do?

Leave the electrodes on, note the symptom and the time on the tracing, and get the provider right away. Do not send the patient back to the waiting room.

3. A patient with lung disease cannot breathe lying flat. How do you record the EKG?

Raise the head only as much as needed, record, and write the position on the tracing, because an upright tracing differs from a flat one.

4. A 10-second rhythm strip shows 9 heartbeats. What heart rate does that suggest?

9 x 6 = 54 beats a minute. Compare it with the patient's pulse and the machine's number, and tell the provider if they disagree.

Key Takeaways

  • State law and your employer decide whether you record EKGs, and you need hands-on training and a documented competency check first.
  • Find the fourth intercostal space by touch from the sternal angle. Never place V1 and V2 by eye.
  • V5 and V6 sit level with V4, and limb electrodes stay on the limbs for a resting EKG.
  • Check speed, gain, all 12 leads, the baseline and lead I before the electrodes come off.
  • Every tracing goes to the provider, and a tracing from a patient with symptoms goes right now.
  • You record the EKG. The provider interprets it.