Running Point-of-Care Tests

15 min read Training Guide

Run CLIA-waived tests the right way by following the maker's instructions exactly, running quality control, collecting a good sample, timing the read and recording the result.

Table of contents

Point-of-care tests are lab tests run right in the clinic, often while the patient waits: urine pregnancy tests, urine dipsticks, rapid strep and flu tests, fingerstick glucose and others. The provider often acts on the result during the same visit, so a wrong result can mean the wrong treatment that day. These tests are simple, but they are not foolproof. One skipped, changed or mistimed step can give a wrong result.

Your state's law and your employer's policy decide whether you may run these tests, and some states add their own lab rules on top of the federal ones. Before you test a patient's sample, you need training on that specific test, practice while an experienced person watches, and a documented competency check. Nothing here certifies you. You run the test and report the result to the provider. The provider decides what it means.

What You Will Be Able to Do

  • Explain what a CLIA-waived test is and what a Certificate of Waiver requires.
  • Run quality control and act on a failed control.
  • Collect and label a sample the way the test requires.
  • Time and read a test, and record the result correctly.
  • Know which results go to the provider right away.

CLIA and Waived Tests

The Clinical Laboratory Improvement Amendments (CLIA) are federal rules, found at 42 CFR Part 493, that cover any site testing human samples to diagnose or treat patients. That includes a small clinic running one urine test. The Centers for Medicare & Medicaid Services (CMS) runs the program, the Food and Drug Administration (FDA) decides which tests are waived, and CDC publishes guidance for testing sites.

A waived test is a simple test with a low risk of a wrong result. Waived tests include ones named in the rule, such as non-automated urine dipsticks and visual urine pregnancy tests, tests the FDA has cleared for home use, and tests the FDA has waived at a maker's request.

A site that runs only waived tests needs a CLIA Certificate of Waiver, generally one for each location. Under 42 CFR 493.15 and 493.35, a site with a Certificate of Waiver must:

  1. Follow the manufacturer's instructions for performing every test.
  2. Allow announced and unannounced inspections, such as after a complaint.

Ask on your first day where the certificate is posted and which tests your clinic runs under it.

Changing a Test Changes Everything

Any change to a waived test makes it an "off-label" use, and CMS and CDC treat an off-label test as high complexity. A site with only a Certificate of Waiver may not run high-complexity tests. It would need a higher-level CLIA certificate and would have to meet far stricter rules.

Off-label use includes:

  • Using a different sample type than the instructions list.
  • Using a swab or collection device that did not come with the kit.
  • Skipping, reordering or shortening any step.
  • Testing a patient outside the limits in the instructions. CDC's example is running a glucose meter on a patient whose blood count (hematocrit) or oxygen level falls outside the range the meter's instructions allow.

If a shortcut seems harmless, it is still off-label. Ask the person who directs testing before you change anything.

Common Tests in a Clinic

Test Sample Where it goes wrong
Urine pregnancy Urine Reading after the time window
Urine dipstick Fresh urine Pads read at the wrong times, or an old sample
Rapid strep Throat swab Swab touched the tongue instead of the tonsils
Rapid flu or COVID-19 antigen Nasal swab Wrong swab type or depth for that kit
Blood glucose Fingerstick Shared lancet device or unclean meter

Before You Test

Read the Current Instructions

The package insert is the rulebook. Keep the current copy where you test. CDC notes that quick-reference cards and color charts help, but they do not replace the instructions. When a new shipment or lot arrives, compare its insert with the one on file. If anything changed, file the old copy and post the new one.

Check Storage and Dates

Store kits, strips and controls at the temperatures printed on the box, and log refrigerator and room temperatures if your clinic does. Some kits must reach room temperature before use. Check the expiration date on every kit and on the control material. Use parts from one kit lot only, and never mix them with parts from another lot.

Run Quality Control

Quality control (QC) proves the test is working before you trust it with a patient.

  • Internal controls are built into the test, such as a control line that shows the test ran and the sample moved through the strip correctly.
  • External controls are samples with known results, usually one positive and one negative, that you test exactly as you would a patient's sample.

Run QC at least as often as the instructions require. CDC also recommends running external controls with each new shipment, each new lot number and each new person who tests. Problems such as a power outage or a hot storage room are also reasons to run QC. Record each run in the QC log, including the date, lot numbers, expiration dates, the expected and actual results, and your initials.

When a Control Fails

Worked example. A new shipment of rapid strep kits arrives with a new lot number. You run the external controls before testing anyone. The negative control shows a test line, so it reads positive. Three patients are waiting.

  1. Stop. Do not run or report patient tests from that lot. CDC says to hold patient results until the problem is found and fixed, for both external and internal control failures.
  2. Look for the cause. Were the steps followed? Are the kits or controls expired? Were they stored correctly? Were the caps switched, contaminating a control?
  3. Fix the problem and repeat QC. If the controls now give the expected results, test the patients.
  4. If QC still fails, try a different lot, and contact the manufacturer or the person who directs testing.
  5. Record the failure and what you did about it.

Tell the provider about the delay so they can plan the visit.

Collecting a Good Sample

  1. Confirm the order, and identify the patient with two identifiers, such as full name and date of birth.
  2. Check any preparation the test needs, such as fasting, and ask whether the patient followed it.
  3. Use only the collection device the kit specifies. CDC notes that swabs are made of different materials, and a substitute can change the result.
  4. Label the sample right away with two identifiers. If the sample goes straight onto a test device, label the device before you collect.
  5. Test within the time the instructions allow, or store the sample exactly as they say.

Throat Swab

Use good light and a tongue depressor. Rub the swab firmly over both tonsils, or where they were, and the back of the throat. Keep it off the tongue, cheeks and teeth, since saliva weakens the sample. Some clinics collect two swabs at once, so a backup culture can be sent if the provider orders one.

Nasal Swab

Each kit says whether it needs a swab from the front of the nose or a deeper one from the back of the nose and throat (nasopharyngeal), how far to insert it and how many times to rotate it. Follow those numbers exactly for that kit.

Urine

For a clean-catch sample, the patient cleans the area, starts to urinate into the toilet, catches the middle of the stream in the cup, and finishes in the toilet. Dipstick tests need fresh urine. For pregnancy tests, first-morning urine is the most concentrated, but follow what the kit accepts.

Fingerstick

Use a single-use lancet that disables itself, and stick the side of the fingertip. Wipe away the first drop if the instructions say to, and do not squeeze hard, since tissue fluid dilutes the blood. CDC says fingerstick devices must never be used on more than one person. A shared glucose meter must be cleaned and disinfected after every use as its maker directs, and a meter whose maker gives no cleaning steps should not be shared.

Warning: Shared fingerstick devices have spread hepatitis B between patients. A lancet holder labeled for one person stays with one person.

Running and Reading the Test

  1. Confirm QC is done and passing for the lot in your hand.
  2. Follow the steps in the exact order the instructions list.
  3. Start a timer the moment the instructions say to.
  4. Read at the stated time. CDC explains that reading too early can give a false negative or an invalid result, because the reaction has not finished. Reading too late can give a false positive from color that keeps developing, a false negative from color that fades, or an invalid result.
  5. Read in good light, and compare against the color chart when the test uses one. On a urine dipstick, each pad has its own read time on the bottle's chart, so time each one.
  6. Treat a missing control line as invalid. Do not report it. Repeat the test with a new device, and a new sample if the instructions require one.

Example of a late read. A kit's instructions say to read at 10 minutes and not after 15. You are pulled into another room and come back at 25 minutes to a faint line. That result cannot be read either way. Run the test again on a new device.

Recording and Reporting Results

CDC's guidance for recording results:

  • Numbers get their units, such as "glucose 142 mg/dL."
  • Yes-or-no results get words, not symbols. Write "Positive" or "Negative," never "+" or "-", which are easy to misread.
  • Invalid results get recorded too. Record the first result, fix the problem, retest, and record the repeat. Report only the final valid result.
  • A result given by phone or in person is documented and followed by a written report.

Also record the test name, date and time, the kit lot number and expiration date (in the chart or the log, as your clinic does), and your initials. Give results only to people authorized to receive them, under HIPAA.

A clean entry looks like this: "Rapid strep antigen, throat swab: Negative. Internal control valid. Collected 10:12, read 10:22. Kit lot and expiration in log. Second swab held for culture per provider. J. Ortiz, MA."

Results That Cannot Wait

Your clinic decides which tests have critical values, results that need the provider's attention right away. Learn that list and how to reach the provider fast, and document when and to whom you reported each one. Some positive results for certain infections must also be reported to public health, and your clinic handles that process.

Common Mistakes

Mistake What it costs
Reading after the time window A false positive or false negative
Using a swab not from the kit Off-label use and an untrustworthy result
Mixing parts from two lots Unreliable results
Testing patients after a failed control Wrong results in the chart
Writing "+" in the chart A misread result

Check Your Understanding

1. To save money, the clinic starts using another brand's swabs with its rapid flu kits. What is the problem?

That is off-label use, which makes the test high complexity, and a Certificate of Waiver does not cover it. Use only the swabs that come with the kit.

2. You run a glucose meter's control solution and get a result outside the range printed on the bottle. Two patients need fingersticks. What do you do?

Do not test patients on that meter yet. Check the expiration and storage of the strips and control, repeat the control, and get help if it fails again. Record what happened.

3. How do you chart a positive urine pregnancy test?

Write "Positive" in words, with the test name, the date and time, the lot information as your clinic records it, and your initials, then tell the provider.

4. A test cassette shows no control line. What does that mean?

The result is invalid. Do not report it. Repeat the test with a new device, and a new sample if the instructions require one.

Key Takeaways

  • State law and your employer decide whether you run these tests, and you need training and a documented competency check on each test first.
  • A Certificate of Waiver requires following the manufacturer's instructions exactly. Any change makes the test high complexity.
  • Run QC as the instructions require, plus with each new shipment, lot and tester, and never report patient results after a failed control.
  • Use only the kit's collection devices, and never share a fingerstick device.
  • Time every read, write results in words with units, and send critical values to the provider right away.